Normal Phase

YMC-Pack Polyamine II
Specification
Particle size : 5µm
Pore size : 120Å
Usable pH range : 2.0-7.5

YMC-Pack Polyamine II is a silica-based packing bonded with polyamine. It is particularly useful for separation of sugars. The column life of YMC-Pack Polyamine II in aqueous mobile phase is longer than conventional silica-based amino columns, and thus is applicable to separation of oligosaccharides using mobile phase with relatively higher water content. In addition, YMC-Pack Polyamine II can be used to separate ionic compounds with a combination of normal-phase mode and weak anion exchange mode.

  • Silica gel chemically bonded to polyamine
  • The most suitable column for separation of sugars
  • Solves durability problem of conventional silica-based amino columns
  • Useful for separation of hydrophilic compounds including vitamins
  • Useful for separation of fat-soluble compounds using nonaqueous mobile phase

Excellent durability

Excellent durability

Flow conditions

Column : YMC-Pack Polyamine II
250 x 4.6mmI.D.
Eluent : acetonitrile/water
Flow Rate : 1.0mL/min
Temperature : ambient
Time : 100 hours

Analytical conditions

Eluent : acetonitrile/water(75/25)
Flow Rate : 1.0mL/min
Temperature : 26°c
Time : RI, 32 x 10-6 RIU/FS

Even in a hard flow durability test using mobile phase that consists only of water,the retention time of sugars changes little.

The most suitable column for separation of sugars

This figure shows differences in selectivity of phenols when separated on Diol-NP and SIL under the same conditions. For these compounds, Diol columns provide longer retention time and better separation.

The most suitable column for separation of sugars

1. Fructose
2. Glucose
3. Sucrose
4. Maltose
5 .Lactose

Column : YMC-Pack Polyamine II, 250 x 4.6mmI.D.
Eluent : acetonitrile/water (75/25)
Flow Rate : 1.0mL/min
Temperature : 26°c
Time : RI, 32 x 10-6 RIU/FS

For normal phase separation
Tocopherol isomers L-ascorbic acid and erythorbic acid
Column : YMC-Pack Polyamine II 250 x 4.6mmI.D.
Eluent : n-hexane/ethyl acetate (70/30)
Flow Rate : 1.0mL/min
Temperature : 26°c
Time : UV at 295 nm
Column : YMC-Pack Polyamine II 250 x 4.6mmI.D.
Eluent : 50mM NH4H2PO4
/acetonitrile (30/70)
Flow Rate : 1.0mL/min
Temperature : 26°c
Time : UV at 250 nm

YMC-Pack Polyamine II is applicable for separation of fat-soluble vitamins and water-soluble vitamins as a normal-phase column that can be used with water or buffer and various organic solvents.