Columns/YMC*Gel

YMC-Pack Diol-SEC
Specification
Particle size :

5µm

Pore size : 60Å,120Å
200Å,300Å
Usable pH range : 5.0-7.5

YMC-Pack Diol-SEC is a size exclusion chromatography column utilizing a silica gel base. Diol-120, 200, and 300 are suitable for separation or molecular weight determination of proteins with molecular weights of 10,000 to several hundred thousands. Diol-60 is the most suitable for separation of peptides of oligosaccharides whose molecular weights are 10,000 or less.

  • Utilizes silica gel base with high resistance to pressure
  • Low-cost gel filtration column
  • Useful for molecular weight determination of proteins and sugars
  • Minimal secondary interactions
Calibration curves of proteins obtained using columns with different pore sizes

Calibration curves of proteins obtained using columns with different pore sizes

1. Glutamate dehydrogenase
2. Lactate dehydrogenase
3. Enolase
4. Adenylate kinase
5. Cytochrome c

 

Column : YMC-Pack Diol-SEC 500 x 8.0 mmI.D.
Eluent : 0.1M KH2PO4-K2HPO4 (pH 7.0)
containing 0.2M NaCl
Flow Rate : 0.7 mL/min
Temperature : ambient
Detection : UV at 280 nm

Diol-120, Diol-200 ,and Diol-300 are suitable for the separation or molecular weight determination of proteins with molecular weights of 10,000 to several hundred thousands.

For separation of proteins with molecular weights of 10,000 to several hundred thousands

For separation of proteins with molecular weights of 10,000 to several hundred thousands

1. Glutamate dehydrogenase
    (MW 290,000)
2. Lactate dehydrogenase
    (MW 142,000)
3. Enolase
    (MW 67,000)
4. Adenylate kinase
    (MW 32,000)
5. Cytochrome c (MW 12,400)

Column : YMC-Pack Diol-SEC 500 x 8.0 mmI.D.
Eluent : 0.1M KH2PO4-K2HPO4 (pH 7.0) containing 0.2M NaCl
Flow Rate : 0.7 mL/min
Temperature : ambient
Detection : UV at 280 nm


For separation of peptides with molecular weights of 10,000 or less

Diol-60

1. Insulin (Bovine)
2. Neurotensin
3. Angiotensin-II
4. Glycine

Column : YMC-Pack Diol-60  500 x 8.0 mmI.D.
Eluent : 0.1M KH2PO4-K2HPO4 (pH 7.0) containing
0.2MNaCl/acetonitrile (70/30)
Flow Rate : 0.7 mL/min
Temperature : ambient
Detection : UV at 215 nm
 
For separation of peptides with molecular weights of 10,000 or less
  MW
1. Myoglobin
2. Ribonuclease A
3. Cytochrome c
4. Insulin (Bovine)
5. Insulin B chain
6. α-Mating factor
7. Neurotensin
8. Angiotensin-I
9. CCK-octapeptide
10. Bradykinin
11. Angiotensin-II
12. Oxytocin
13. Angiotensin-III
14. Met-enkephalin
15. Leu-enkephalin
16. Tetraglycine
17. Glycine
17,000
13,700
12,400
5,700
3,496
1,684
1,672
1,296
1,143
1,060
1,040
1,007
931
573
555
246
75

For molecular weight determination of sugars


Diol-200
For molecular weight determination of sugars

For separation or molecular weight determination of neutral sugars and water-soluble high-molecular-weight compounds, Diol-60, Diol-120, Diol-200,and Diol-300 are useful individually or in combination.